What this guide covers
What counts as a legacy device
A legacy device is one CE marked under the old Medical Devices Directive, the Active Implantable Medical Devices Directive or the IVD Directive, which is still being placed on the EU market under the MDR's transition rules. Most have certificates that were extended under Regulation (EU) 2023/607, which gave manufacturers more time to move to the MDR. The device must not have changed significantly in design or intended purpose, and the manufacturer must be working towards MDR certification.
If your device has already been certified under the MDR and registered as an MDR device, it isn't a legacy device for this purpose.
Where the deadline comes from
Regulation (EU) 2024/1860 changed how EUDAMED is introduced, so that each module can become mandatory once it has been shown to work, instead of waiting for the whole system. When the Commission published its decision on 27 November 2025 confirming that the first four modules worked, two clocks started. New devices had to be registered from 28 May 2026. Legacy devices still on the market got twelve months, which ends on 28 November 2026.
The main exceptions
Not every legacy device needs its own registration:
- If an MDR version of the same device, with the same identification, risk class and characteristics, is already registered in EUDAMED, the legacy device is linked to it rather than registered separately.
- If a legacy device is no longer placed on the market by the deadline, it doesn't need registering. That changes if a serious incident or field safety corrective action involving it is reported later. At that point it has to be registered so the report can refer to it.
- Custom-made devices are outside EUDAMED device registration altogether.
The second point is the one that catches people out. A device you stopped selling two years ago can still need a registration if something goes wrong with units in the field.
The identifiers EUDAMED gives you
Devices certified under the old directives don't have a Basic UDI-DI, which EUDAMED normally uses to group a device family. So EUDAMED generates two stand-in identifiers for legacy devices: a EUDAMED-DI, which takes the place of the Basic UDI-DI, and a EUDAMED-ID, which takes the place of the UDI-DI. If you already have a UDI-DI for the device, you can enter it and EUDAMED generates the EUDAMED-DI alongside it.
These identifiers belong to the legacy registration. When the device moves to the MDR, it gets a proper Basic UDI-DI and UDI-DI, and the legacy identifiers aren't carried over.
A checklist for the next few weeks
- List every device you still place on the EU market under a directive certificate.
- For each one, check whether an MDR version is already registered. If it is, confirm the link rather than creating a new entry.
- Make sure your company's actor registration and SRN are in order, along with those of your authorised representative if you need one.
- Gather the certificate details, risk class and any existing UDI-DIs before you start entering data.
- Keep a list of legacy devices you've stopped selling, so you know which ones would need registering if an incident is reported.
- Check afterwards that the public record matches your technical documentation. Your importers and distributors will be looking at it.
After the deadline
Once legacy devices are in, EUDAMED will hold the most complete list of devices on the EU market there has been. That's useful well beyond compliance. Our guide to using EUDAMED's public data covers how post-market teams can use it.
MedFortis shows a device's current EUDAMED registration status alongside its recalls, safety notices and adverse events, so once your registrations are in, you can check them next to everything else regulators have published about the device.
Check your devices in one search
Book a free 30-minute demo. We'll look up your device lines live and show their EUDAMED status next to their regulatory history.
Book a Demo →Key takeaways
- Legacy devices are devices still on the EU market under directive certificates during the MDR transition.
- They must be registered in EUDAMED by 28 November 2026, twelve months after the Commission's decision of 27 November 2025.
- You don't need a separate entry if a matching MDR device is already registered, or if the device is off the market, unless an incident is reported later.
- EUDAMED assigns a EUDAMED-DI and EUDAMED-ID in place of the UDI identifiers.